| Catalog Number |
0035280 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the user found foreign material (like a black ink stain) on the bulb irrigation syringe prior to use.Per additional information from the ibc via email 01-mar-2020, there was a black material or mark just between the seal of where the bulb and syringe came together.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed since the device did not fail to meet relevant specifications.Visual evaluation of the returned sample noted one opened (with original packaging), bulb irrigation syringe.Visual inspection of the sample noted multiple (5) brown/black spots between the syringe body and bulb measuring an aggregate 0.30 sq mm.This is within specification per inspection procedure which states, "loose foreign matter or embedded shall not exceed an aggregate total of 0.6mm² or 1/16¿ per the dirt estimation chart." a device history record review was not required per the investigation.The instructions for use were found adequate and state the following: "this is a single use device.Do not re-sterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the user found foreign material (like a black ink stain) on the bulb irrigation syringe prior to use.Per additional information from the ibc via email 01mar2020, there was a black material or mark just between the seal of where the bulb and syringe came together.
|
| |
|
Search Alerts/Recalls
|