• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US LUPINE STABILIZATION SAWTOOTH DRGDE *EA; SURGICAL HAND DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US LUPINE STABILIZATION SAWTOOTH DRGDE *EA; SURGICAL HAND DRILL Back to Search Results
Model Number 213819
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/05/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The lot number is unknown.
 
Event or Problem Description
It was reported by the sales rep via phone that during a labral shoulder repair the teeth of two sawtooth drill guides were bent.Another device was used to complete the procedure.No patient consequences or surgical delay reported.The devices were discarded by the hospital.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that there was no lot number available.It was further reported that the tooth was bent from normal wear and tear.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.One possible root cause of the reported problem could be normal wear and tear.However without the return of the complaint device, and no further information provided we cannot determine a definitive root cause.Further, no lot numbers were supplied which precludes conducting a dhr review or a lot specific search in the complaints handling system.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
SURGICAL HAND DRILL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9867794
Report Number1221934-2020-00957
Device Sequence Number19593426
Product Code FZX
UDI-Device Identifier10886705003303
UDI-Public10886705003303
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number213819
Device Catalogue Number213819
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/09/2020
Supplement Date Received by Manufacturer03/21/2020
03/31/2020
Initial Report FDA Received Date03/23/2020
Supplement Report FDA Received Date03/30/2020
04/01/2020
Patient Sequence Number1
-
-