| Model Number |
0167L16 |
| Medical Device Problem Code |
Misassembled (1398)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330)
|
| Date of Event |
02/28/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that there were several faulty catheters from the same batch.The eyelets from one of the catheters was not uniformed, the balloon from one catheters inflated on one side, and one of the catheters was found kinked/bent upon removal.All of the reported events allegedly caused unnecessary discomfort and pain to the patients.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive due to a poor sample and incorrect sample.A photo of an eggshell white catheter was returned showing two eyes.0167l16 is an amber three way irrigation catheter with three eyes (2 drainage and 1 irrigation).The drainage eyes are intended to be staggered.As the returned catheter contains is a different color than 0167l16 and all sides of the catheter is not shown, the reported event cannot be confirmed.A potential root cause could be due to an "operator error" during the eye burning process.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.Or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.For urological use only to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water." h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that there were several faulty catheters from the same batch.The eyelets from one of the catheters was not uniformed, the balloon from one catheters inflated on one side, and one of the catheters was found kinked/bent upon removal.All of the reported events allegedly caused unnecessary discomfort and pain to the patients.
|
| |
|
Search Alerts/Recalls
|