MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VENAFLO VASCULAR GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VENAFLO VASCULAR GRAFT Back to Search Results
Model Number VLT4046C
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 02/25/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.(expiry date: 10/2024).
 
Event or Problem Description
It was reported that post-surgical graft implantation in avf, the surgical graft leaked fluid and the patient experienced infection within the hospital.It was further reported that the infection was treated for two weeks with antibiotics and that the surgical graft will remain implanted.The healthcare provider (hcp) will continue to observe the patient's symptoms.The patient is reportedly stable.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.The lot met all release criteria.Investigation summary: a review of the sterilization records and bioburden identified that the lot met all sterile release criteria and no unacceptable bioburden results were identified.A lot sample was returned and tested using the wep tester at 3.5psi.No water was observed to be leaking from the device.Based on the available information, the most likely root cause is patient and/or procedural related.The investigation is inconclusive.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that post-surgical graft implantation in avf, the surgical graft leaked fluid and the patient experienced infection within the hospital.It was further reported that the infection was treated for two weeks with antibiotics and that the surgical graft will remain implanted.The healthcare provider (hcp) will continue to observe the patient's symptoms.The patient is reportedly stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENAFLO VASCULAR GRAFT
Common Device Name
VASCULAR GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9869368
Report Number2020394-2020-01964
Device Sequence Number19552825
Product Code DSY
UDI-Device Identifier00801741020582
UDI-Public(01)00801741020582
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K052282
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVLT4046C
Device Catalogue NumberVLT4046C
Device Lot NumberVTDX0456
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Initial Date Received by Manufacturer 02/25/2020
Supplement Date Received by Manufacturer06/03/2020
Initial Report FDA Received Date03/23/2020
Supplement Report FDA Received Date06/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CEFTRIAXONE; CEFTRIAXONE; GENTAMICINE; GENTAMICINE
Outcome Attributed to Adverse Event Other;
-
-