| Model Number |
VLT4046C |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number for the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.(expiry date: 10/2024).
|
| |
|
Event or Problem Description
|
|
It was reported that post-surgical graft implantation in avf, the surgical graft leaked fluid and the patient experienced infection within the hospital.It was further reported that the infection was treated for two weeks with antibiotics and that the surgical graft will remain implanted.The healthcare provider (hcp) will continue to observe the patient's symptoms.The patient is reportedly stable.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.The lot met all release criteria.Investigation summary: a review of the sterilization records and bioburden identified that the lot met all sterile release criteria and no unacceptable bioburden results were identified.A lot sample was returned and tested using the wep tester at 3.5psi.No water was observed to be leaking from the device.Based on the available information, the most likely root cause is patient and/or procedural related.The investigation is inconclusive.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
|
| |
|
Event or Problem Description
|
|
It was reported that post-surgical graft implantation in avf, the surgical graft leaked fluid and the patient experienced infection within the hospital.It was further reported that the infection was treated for two weeks with antibiotics and that the surgical graft will remain implanted.The healthcare provider (hcp) will continue to observe the patient's symptoms.The patient is reportedly stable.
|
| |
|