• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR EAST; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION CAPIOX RX25 OXYGENATOR EAST; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*RX25RE
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The subsidiary reported to terumo cardiovascular that during quality inspection, they noted the oxygenator had no label.No patient involvement.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on march 23, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date); g4 (date received by manufacturer); g7 (indication that this is a follow-up report); h2 (follow-up due to additional information); h4 (device manufacture date); h6 (identification of evaluation codes 11, 3331, 4114, 150, 25).Method code #1:11 - testing of device from same lot/batch retained by manufacturer.Method code #2: 3331 - analysis of production records.Method code #3: 4114 - device not returned.Results code: 150 - labeling and instructions for use/maintenance.Conclusions code: 4315 - cause not established.The sample was not returned for investigation, however the photos provided with complaint confirmed the missing box label.The retention sample was inspected and was found to be labeled properly.Review of device history were reviewed indicating no abnormalities related to the issue.Capiox labeling is 100% inspected and label reconciliation is completed ensure each part meets tcvs specifications.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPIOX RX25 OXYGENATOR EAST
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key9869552
Report Number1124841-2020-00073
Device Sequence Number13602143
Product Code DTZ
UDI-Device Identifier00699753450103
UDI-Public(01)00699753450103
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K153213
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/30/2022
Device Model Number3CX*RX25RE
Device Catalogue NumberN/A
Device Lot NumberXP09
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/02/2020
Supplement Date Received by Manufacturer04/29/2020
Initial Report FDA Received Date03/23/2020
Supplement Report FDA Received Date05/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-