• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COCHLEAR BONE ANCHORED SOLUTIONS AB FLANGE FIXTURE ST 4MM W. BAHA ABUTMENT COCHLEAR BAHA CONNECT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COCHLEAR BONE ANCHORED SOLUTIONS AB FLANGE FIXTURE ST 4MM W. BAHA ABUTMENT COCHLEAR BAHA CONNECT SYSTEM Back to Search Results
Model Number 90434
Device Problem Loss of Osseointegration (2408)
Patient Problem Patient Problem/Medical Problem (2688)
Event Type  Injury  
Manufacturer Narrative
The reported adverse is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time. Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication. Additionally, there are no indications that a product failure has contributed to the reported issue. Implant loss with baha implants can occur at any time following implantation. Known reasons for implant loss include lack of adequate bone quality/quantity, trauma, infection, generalised diseases and surgical complication. The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis. This event is added to this monitoring. This report is submitted on march 24, 2020.
 
Event Description
Per the clinic, it was reported that the patient experienced a loss of osseointegration resulting in fixture loss when the patient removed their external equipment (date not reported). It is unknown whether there are plans to reimplant the patient due to potential bone quality issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLANGE FIXTURE ST 4MM W. BAHA ABUTMENT
Type of DeviceCOCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
po box 82
mölnlycke, 435 2 2
SW 435 22
Manufacturer Contact
bianca pries
1 university avenue
macquarie university, nsw 2109
AS   2109
MDR Report Key9869734
MDR Text Key184637713
Report Number6000034-2020-00787
Device Sequence Number1
Product Code LXB
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
NI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number90434
Device Catalogue Number90434
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/26/2020
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 03/23/2020 Patient Sequence Number: 1
-
-