MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; 3-WAY BARDEX FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER; 3-WAY BARDEX FOLEY CATHETER Back to Search Results
Model Number 0167L16
Medical Device Problem Code Dent in Material (2526)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 02/28/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that there were several faulty catheters from the same batch: the eyelets from one of the catheters were not uniformed, the balloon from one catheter inflated on one side, and one of the catheters was found kinked/bent upon removal.Reportedly, all of the events caused unnecessary discomfort and pain to the patients.
 
Additional Manufacturer Narrative
The reported event was confirmed.A photo sample of an amber lubricath catheter was returned with an asymmetric balloon.However as it is unknown whether the device was correctly inflated the cause of the issue could not be confirmed.A potential root cause could be due to "balloon molding".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Valve type: use luer slip tip syringe.Do not use needle.Recommended inflation capacities 3cc balloon: use 5ml sterile water 5cc balloon: use 10ml sterile water 15cc balloon: use 20ml sterile water 20cc balloon: use 25ml sterile water 30cc balloon: use 35ml sterile water 40cc balloon: use 45ml sterile water 75cc balloon: use 80ml sterile water".
 
Event or Problem Description
It was reported that there were several faulty catheters from the same batch: the eyelets from one of the catheters were not uniformed, the balloon from one catheter inflated on one side, and one of the catheters was found kinked/bent upon removal.Reportedly, all of the events caused unnecessary discomfort and pain to the patients.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER
Common Device Name
3-WAY BARDEX FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9871078
Report Number1018233-2020-02035
Device Sequence Number2377869
Product Code KOD
UDI-Device Identifier00801741018947
UDI-Public(01)00801741018947
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model Number0167L16
Device Catalogue Number0167L16
Device Lot NumberNGCZ2156
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2020
Initial Date Received by Manufacturer 03/06/2020
Supplement Date Received by Manufacturer06/02/2020
Initial Report FDA Received Date03/24/2020
Supplement Report FDA Received Date06/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-