• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LCS COMP RP INSERT LG+ 10MM; LCS COMPLETE : KNEE TIBIAL INSERT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LCS COMP RP INSERT LG+ 10MM; LCS COMPLETE : KNEE TIBIAL INSERT Back to Search Results
Model Number 1294-05-710
Medical Device Problem Codes Fracture (1260); Naturally Worn (2988)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
Date of Event 03/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Revision of lcs implanted in 2011 by (b)(6).Patient reported pain and difficulty mobilising, x-ray revealed tibial peri-prosthetic fracture & osteolysis on the lateral femoral condyle.On removal of the implants, marked delamination and a large crack on the medial side of the polyethylene.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.Corrective action was not indicated.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified that could have attributed to the event.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCS COMP RP INSERT LG+ 10MM
Common Device Name
LCS COMPLETE : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9871522
Report Number1818910-2020-09003
Device Sequence Number2910566
Product Code NJL
UDI-Device Identifier10603295022220
UDI-Public10603295022220
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1294-05-710
Device Catalogue Number129405710
Device Lot NumberE5EDB4
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/06/2020
Supplement Date Received by Manufacturer06/03/2020
Initial Report FDA Received Date03/24/2020
Supplement Report FDA Received Date06/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
COMPETITOR CEMENT; COMPETITOR CEMENT; LCS COMPLETE FEM POR R LG+; LCS COMPLETE FEM POR R LG+; MBT CEM TIBIAL TRAY SZ5; MBT CEM TIBIAL TRAY SZ5
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-