| Model Number |
PCU060300130 |
| Medical Device Problem Codes |
Inflation Problem (1310); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Image review: two photographic images were received for evaluation.The first image is of a procedural cine image of the patient¿s mid right superficial femoral artery based on anatomical landmarks.In the cine image the pta balloon catheter is inflated and three balloon twists ¿candy wrapping¿ are visible.The second image the pta balloon catheter has been removed from the patient and is inflated.In this photograph at the proximal end of the balloon chamber there are only two balloon twists ¿candy wrapping¿.It is noted that the inner guidewire lumen is undulating within the balloon chamber an indication that the balloon may have been over inflated.If information is provided in the future, a supplemental report will be issued.
|
| |
|