| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer's report was duplicated and attributed to a faulty integrated circuit on the analog board.The analog board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to monitor a patient (age & gender unknown), the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|