Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Date 01/31/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-01280.
|
|
Event Description
|
It was reported the e1 liner would not assemble with g7 acetabular shell.The acetabular bone fractured when the liner was driven into the shell.This surgery was finished with a backup liner.It was reported that no further information is available.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
Complaint sample was evaluated and the reported event was confirmed.Upon visual inspection there is damage to the locking feature and circular indentations on the outside diameter of the device.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|