SMITH & NEPHEW ORTHOPAEDICS LTD ACETABLR CUP HAP 44MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
| Model Number |
74120144 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
12/27/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a revision surgery on the right hip was performed due to pain, failure of hardware, elevated levels of cobalt and chromium, metal toxicity, metallosis, pseudotumors, bone and tissue loss, and loss of mobility.The patient was previously implanted with a bhr resurfacing system.There is no information regarding to the devices explanted on the revision surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that right hip revision surgery was performed.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the head and cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the cup.Similar complaints have been identified for the head.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All the released devices involved met manufacturing specifications at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.A risk management review was performed.No additional risks were identified as result of the reported event.The available medical documents were reviewed.The reported pain, elevated metal ion levels and noted intraoperative findings of blackened pseudotumor may be consistent with findings associated with metal debris; however, the root cause of the pain, elevated metal ions and pseudotumor cannot be confirmed and it cannot be concluded that the reported reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
|
| |
|
Search Alerts/Recalls
|
|
|