|
Pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.No unlocked j2/lcd keypad connector.Perform brush cleaning with the isopropyl alcohol and cleaning up all corrosion on the keypad.Unit received with normal operating current.The stop (idle) current and run current measurement tests are within specification.Pump passed displacement test and a21 alarm test.No battery power loss alarm during testing.Pump history download using thds was successful.However, a47 alarms found on the history file which caused corrupted data.Unable to confirm the reported event date due to the corrupted data.A47 alarms are due to the pump not having power for a long period of time.The following were noted during visual inspection: cracked belt clip slot, cracked case at the reservoir tube window corners, cracked reservoir tube window, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.The test p-cap/reservoir does lock into place.No unexpected battery power loss anomalies noted.Battery out limit alarm not confirmed.No button error alarm during testing.However, intermittent button response was confirmed due to corroded keypad traces.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|