| Model Number |
MT22719 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that intermittent audio output occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An external visual inspection was performed and passed.Charge and boot tests were performed and passed.Functional tests were performed and passed all relevant testing.Manual functional "try it" tests were performed and passed.Speaker audio wiggle tests were performed and passed.An internal visual inspection was performed and passed.The speaker resistance was measured and found to be within specification.The receiver log was downloaded for review, finding no error related to the complaint.The allegation was not confirmed.A probable cause could not be determined as the allegation was not found.No injury or medical intervention was reported.
|
| |
|