|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 09/2021).
|
|
It was reported that approximately one year two months post port device implant, the catheter allegedly broke at the connection point and migrated to the heart, which was confirmed by x-ray imaging.Reportedly, the catheter was removed surgically by interventional radiology.The patient is reported as stable.
|