| Model Number |
TRF-180 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Capsular Contracture (1761); Seroma (2069)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The events of "capsular contracture and seroma" are physiological complications and analysis of the device generally does not assist allergan in determining a probable cause for these events.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown, "bilateral fluid" and patient¿s concern with the product.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported exchange from textured to smooth implants due to concern with the product.Patient reported left side capsular contracture baker grade unknown and "fluid".Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.Visual evaluation: visual analysis of the photographs identified: biological tissue and red particles on the shell.Device analysis performed through photographs, due to the impossibility to perform microscopic analysis it is not possible to determine the most likely failure mode.A review of the device history record has been completed.No deviations or non-conformances noted.Additional, changed, and/or corrected data: b.5., b.6., d.7., g.3., h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|