• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERLOC SAFETY INFUSION SET 20G X 1.5IN; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERLOC SAFETY INFUSION SET 20G X 1.5IN; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Codes Exposure to Body Fluids (1745); Needle Stick/Puncture (2462)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asdns0074 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (asdns0074) have been reported from facility.- attachment: [(b)(4)].
 
Event or Problem Description
It was reported via medwatch "powerloc used for access to port.The "lock out" legs did not engage on retraction leaving the needle exposed which caused a needle stick to provider.The stick came just as the needle was being placed in sharps container so the actual needle is not available.We immediately went to the same supply bin and tested another one from what we believe is the exact same lot and the defect was duplicated.We pulled that lot and have a sample from that lot available for analysis." this report addresses the second of two devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 20G X 1.5IN
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key9878951
Report Number3006260740-2020-01017
Device Sequence Number2877885
Product Code FPA
UDI-Device Identifier00801741047756
UDI-Public(01)00801741047756
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0652015
Device Lot NumberASDNS0074
Was Device Available for Evaluation? No
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2020
Initial Report FDA Received Date03/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-