| Model Number |
N/A |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Codes |
Exposure to Body Fluids (1745); Needle Stick/Puncture (2462)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asdns0074 showed one other similar product complaint(s) from this lot number.The complaints for this lot number (asdns0074) have been reported from facility.- attachment: [(b)(4)].
|
| |
|
Event or Problem Description
|
|
It was reported via medwatch "powerloc used for access to port.The "lock out" legs did not engage on retraction leaving the needle exposed which caused a needle stick to provider.The stick came just as the needle was being placed in sharps container so the actual needle is not available.We immediately went to the same supply bin and tested another one from what we believe is the exact same lot and the defect was duplicated.We pulled that lot and have a sample from that lot available for analysis." this report addresses the second of two devices.
|
| |
|