|
BARD ACCESS SYSTEMS POWERLOC MAX POWER-INJECTIBLE INFUSION SET 22G X 0.75IN; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged infusion set was confirmed.The product returned for evaluation was one 22ga x 0.75¿ powerloc max safety infusion set.Usage residues were observed throughout the sample and the safety mechanism was engaged.A needleless injection cap was attached to the luer adapter.A partially circumferential split was observed at the luer/tubing joint.Microscopic inspection of the split revealed a granular fracture surface.Beach marks were observed throughout the fracture surface.Material buckling and discoloration were observed in the vicinity of the split.The split characteristics and material buckling were consistent with material failure due to repetitive stress.It appeared that the tubing fractured, in part, due to repetitive torsional (twisting) stress, suggesting that securement, access and maintenance techniques may have contributed; however, early signal detection system indicates an additional investigation is necessary.Both bd and the supplier are currently investigating corrective actions to mitigate this type of incident and the complaint sample was manufactured prior to the implementation of any such actions.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing on the port access needle cracked.No other information was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing on the port access needle cracked.No other information was provided.(b)(6) 2020 - per air: "patient port tubing distal to the needle but above the cap had a small hole.When [complainant] went to draw labs, upon flushing normal saline, i discovered that saline was squirting out of the hole.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged infusion set was confirmed.The product returned for evaluation was one 22ga x 0.75¿ powerloc max safety infusion set.Usage residues were observed throughout the sample and the safety mechanism was engaged.A needleless injection cap was attached to the luer adapter.A partially circumferential split was observed at the luer/tubing joint.Microscopic inspection of the split revealed a granular fracture surface.Beach marks were observed throughout the fracture surface.Material buckling and discoloration were observed in the vicinity of the split.The split characteristics and material buckling were consistent with material failure due to repetitive stress.It appeared that the tubing fractured, in part, due to repetitive torsional (twisting) stress, suggesting that securement, access and maintenance techniques may have contributed; however, early signal detection system indicates an additional investigation is necessary.Both bd and the supplier are currently investigating corrective actions to mitigate this type of incident and the complaint sample was manufactured prior to the implementation of any such actions.A lot history review (lhr) review is not possible; the manufacturing lot number that was provided is invalid.
|
| |
|
Additional Manufacturer Narrative
|
|
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a damaged infusion set was confirmed.The product returned for evaluation was one 22ga x 0.75¿ powerloc max safety infusion set.Usage residues were observed throughout the sample and the safety mechanism was engaged.A needleless injection cap was attached to the luer adapter.A partially circumferential split was observed at the luer/tubing joint.Microscopic inspection of the split revealed a granular fracture surface.Beach marks were observed throughout the fracture surface.Material buckling and discoloration were observed in the vicinity of the split.The split characteristics and material buckling were consistent with material failure due to repetitive stress.It appeared that the tubing fractured, in part, due to repetitive torsional (twisting) stress, suggesting that securement, access and maintenance techniques may have contributed; however, early signal detection system indicates an additional investigation is necessary.Both bd and the supplier are currently investigating corrective actions to mitigate this type of incident and the complaint sample was manufactured prior to the implementation of any such actions.A lot history review (lhr) review is not possible; the manufacturing lot number that was provided is invalid.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing on the port access needle cracked.No other information was provided.(b)(6) 2020 - per air: "patient port tubing distal to the needle but above the cap had a small hole.When [complainant] went to draw labs, upon flushing normal saline, i discovered that saline was squirting out of the hole.".
|
| |
|
Search Alerts/Recalls
|
|
|