ST. JUDE MEDICAL, COSTA RICA LTDA INQUIRY STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81174
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 01/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
One decapolar, inquiry steerable diagnostic catheter was received for evaluation.The device was returned due to a display issue.Electrode 4 read as an open circuit.A biological foreign material was noted covering electrode 4, interfering with the conductivity of the electrode and consistent with the open circuit detected and the reported display issue.Electrodes 1-3 and 5-10 met specifications of acceptable resistance values.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the electrical issue is due to the biological foreign material covering the electrode.
 
Event or Problem Description
This report is to advise of an event observed during analysis, confirming tissue on the returned device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY STEERABLE DIAGNOSTIC CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key9880390
Report Number3008452825-2020-00164
Device Sequence Number3183418
Product Code DRF
UDI-Device Identifier05414734303361
UDI-Public05414734303361
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K961924
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model NumberIBI-81174
Device Catalogue NumberIBI-81174
Device Lot Number7211538
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2020
Initial Report FDA Received Date03/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-