| Catalog Number |
713.30.05 |
| Medical Device Problem Code |
Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
02/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The instrument set available to the surgeon during the procedure had been incorrectly prepared, substituting the sharp-tipped forceps for a blunt-tipped reposition forceps of the same size that the surgeon was accustomed to using.
|
| |
|
Event or Problem Description
|
|
During a sternal fixation procedure, a bone holding forceps with sharp tips was mistakenly used instead of a blunt-tipped reposition forceps, causing bleeding of the manubrium and requiring bovie cauterization.Ultimately the procedure was completed as planned.
|
| |
|
Additional Manufacturer Narrative
|
|
Fields d1, d2 and d4 have been amended due to corrected information received from the initial reporter.
|
| |
|
Additional Manufacturer Narrative
|
|
Fields d2 and d3 have been amended.
|
| |
|