• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS MIDAS REX® LEGEND FOOTED ATTACHMENT; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS MIDAS REX® LEGEND FOOTED ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AF02-R
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.Evaluation determined that a portion of the foot of the attachment was perforated.It was also noted that the bearings were corroded, the color band and laser markings were faded and illegible.Previous investigation performed under (b)(4) determined that the likely causes are debris in the collet and improper insertion of the tool.The user manual contains the following warnings ¿do not use a legend attachment if any part of the attachment appears to be bent, loose, missing or damaged.Do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff." in addition, ¿insert dissecting tool into motor collet with a slight rotational motion.A tactile and audible click is observed indicating that the dissecting tool is fully seated.¿ the preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use with no record of factory service during this period.We will continue to track and trend this complaint type.(b)(6).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of attachment foot cut by tool.No patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDAS REX® LEGEND FOOTED ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key9882496
Report Number1625507-2020-00036
Device Sequence Number14089861
Product Code HBB
UDI-Device Identifier00613994422224
UDI-Public00613994422224
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAF02-R
Device Catalogue NumberAF02-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2020
Initial Report FDA Received Date03/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-