| Model Number |
1536549 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that whist performing an adapt to shape position (ats) it was impossible to get a calculation (calculation failure).The customer was able to calculate once they used adapt to position (atp). the customer also reported that the motion monitoring structure was displaced.No patients were mistreated.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 updated.H10 updated: it was determined that tsm is causing a contour position shift relative to the images.This can occur for both single and multiple contours (for example, multiple metastatic contours or bifurcated contours).When this occurs, the position of the delineated contours with respect to the anatomy that is displayed will be erroneous.Users could make an incorrect assessment of the monitored volume position relative to the beam during treatment.This could result in a geometric miss resulting in a severity of serious mistreatment.Based on available information there was no mistreatment.Elekta recall is fca-el-0011.An important field safety notice (200-01-801-010) was released in august 2020.An ifsm is estimated to be released april 2021.
|
| |
|