Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to user error.Udi: (b)(4).
|
|
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device had corrosion/rusting/pitting, heat, worn bearings and fluid damage.It was further determined that the device failed pretest for temperature assessment.It was noted in the service order that the device burr was not detachable.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|