| Model Number |
7818 |
| Medical Device Problem Code |
Missing Test Results (3267)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Due to the covid-19 pandemic, the investigation for this event will be delayed.We will however, send the follow-up report as soon as our laboratory services are able to return to normal.The patient/family was the initial reporter, so personal information was not entered.No information was captured as the customer's weight was not provided.
|
| |
|
Event or Problem Description
|
|
The customer reported that his blood glucose readings were not being stored in the memory of the contour next one meter.There was no allegation of an adverse event.The customer was advised to return the device for evaluation.Replacement meter and test strips were sent to the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer returned the suspected contour next one meter and contour next test strips for evaluation.The returned meter was tested with the returned test strips, which gave a satisfactory performance.All blood glucose readings obtained during in-house testing were properly stored in the meter's memory.Additionally, the device history records were reviewed for the suspected contour next one meter and contour next test strips, and no manufacturing anomalies were found.
|
| |
|