Model Number N/A
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not returned by the hospital.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Concomitant medical products: 42522000510, articular surface, lot # 64295096, 42502806402, femoral component, lot # 64238042, 42540200035, patella, lot # 64021230.Report source: (b)(6).
 
Event or Problem Description
It was reported that approximately 7 months post implantation, the patient was revised due to subsidence of the tibial component.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Photographs of the removed implants showed some bone growth on the base porous side, confirming the revision.X-ray review by a hcp confirmed that there was medial tray subsidence with abnormal lucencies along the tibial implant with loosening.Bone quality was osteopenic.Review of the device history records did not identify any deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURAL TIBIA TRABECULAR METAL TWO-PEG POROUS FIXED BEARING RIGHT SIZE F
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9886243
Report Number0001822565-2020-01128
Device Sequence Number2897221
Product Code MBH
UDI-Device Identifier00889024510586
UDI-Public(01)00889024510586
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K173417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42530007502
Device Lot Number64055801
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/13/2020
Supplement Date Received by Manufacturer04/24/2020
Initial Report FDA Received Date03/26/2020
Supplement Report FDA Received Date05/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight60
-
-