|
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿unable to use syringe to inflate the foley balloon¿ with a potential root cause of ¿inflation water is not present in syringe (syringe barrel is cracked or tip cap is missing)¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "directions for use: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.1.Open csr wrap to form sterile field.2.Place underpad beneath patient, plastic side down.3.Put on cuffed gloves.4.Position drape on patient.5.Open lubricant/lubricate catheter.6.Open packet of swabs.7.Prep patient with swabs.8.Proceed with catheterization in usual manner.To inflate catheter, simply insert tip of water-filled syringe gently into valve (do not overpenetrate) and depress plunger.Instill entire amount of sterile water.9.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.10.If specimen is required, fill sterile container from catheter.11.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|