Model Number B11LT
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # t92p0r.Date of event is unknown.Investigation summary: the analysis results found that the b11lt device was received with the tip of the sleeve melted.The obturator was returned with no apparent damage.The event described could not be confirmed as the clear lens was noted without damage.One possible cause for this type of damage may be interaction with an energized device used during the procedure.Caution should be taken to avoid contact between an energized device and the trocar during the surgical procedure.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an unknown procedure, the product showed factory damage as the tip was broken.It is unknown how the procedure was completed.Patient consequence was not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOPATH XCEL BLADELESS TROCAR
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key9886421
Report Number3005075853-2020-01812
Device Sequence Number776678
Product Code GCJ
UDI-Device Identifier10705036001126
UDI-Public10705036001126
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/29/2024
Device Model NumberB11LT
Device Catalogue NumberB11LT
Device Lot NumberT4047N
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/20/2020
Initial Report FDA Received Date03/26/2020
Patient Sequence Number1
-
-