(b)(4).Batch # t92p0r.Date of event is unknown.Investigation summary: the analysis results found that the b11lt device was received with the tip of the sleeve melted.The obturator was returned with no apparent damage.The event described could not be confirmed as the clear lens was noted without damage.One possible cause for this type of damage may be interaction with an energized device used during the procedure.Caution should be taken to avoid contact between an energized device and the trocar during the surgical procedure.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|