Model Number CUSTOM 4.0 FLEXTEND TTS NS +1/EA
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Laceration(s) (1946); Respiratory Distress (2045); Injury (2348)
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received that smiths bivona® adult tts¿ tracheostomy tracheostomy|silicone - bivona tubes neo/ped flextend plus a child managed to pull out a wire on the trach tube causing the tube to split and emergent trach change out.Reported the child hurt a finger when this incident occurred.Photos show damage.Trach change out resolved issue.
 
Additional Manufacturer Narrative
Evaluation results: one bivona tracheostomy tube was returned for investigation in used condition.Visual inspection was performed at 12 inches and under normal conditions of illumination so as to look for any damages to the parts.The visual inspection revealed that the shaft was broken and that the wire was exposed.There was also a hole in the adhesive that was attached to the connector.The investigator also examined photographs submitted by the customer.In the first picture, a cut was clearly visible in the middle of the tube.In the second picture, the broken connector and an exposed wire were visible.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.The above product problems most likely occurred after the device left the manufacturing facility.A definitive root cause was not established however.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA ADULT TTS TRACHEOSTOMY TUBE
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD,INC.
6000 nathan lane north
minneapolis, mn
MDR Report Key9887056
Report Number3012307300-2020-02259
Device Sequence Number16973785
Product Code JOH
UDI-Device Identifier15021312516821
UDI-Public15021312516821
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K944178
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberCUSTOM 4.0 FLEXTEND TTS NS +1/EA
Device Catalogue Number60PFP60
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2020
Supplement Date Received by Manufacturer07/07/2020
Initial Report FDA Received Date03/26/2020
Supplement Report FDA Received Date08/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-