| Model Number |
MMT-723RNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
03/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received multiple motor errors and no delivery alarm.The customer also reported that they were able to clear the alarm.The customer was able to complete insulin pump rewind.The customer reported that the drive support cap was normal.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No motor error alarm or unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and displacement test.The motor was tested outside the device and passed.(b)(4).
|
| |
|