The report of the autopulse platform (sn (b)(6)) displaying a system error, out of service, revert to manual cpr message was not confirmed during functional testing and archive data review.The platform performed as intended.No device malfunction.Per archive data, the last time the autopulse platform was used or powered on by customer on (b)(6) 2019 and no system error was recorded.Per customer, the problem occurred date was on (b)(6) 2020 and there was no evidence on the archive data showing that the platform was used on the customer reported event date.Probably, the customer reported on a wrong platform.Unrelated to the reported complaint, cracked top cover and cracked short black cover was observed during visual inspection.Root cause were likely attributed to mishandling such as a drop.During functional testing, no issues were observed.During archive data review, no issues were observed.Awaiting customer's approval for service repair.
|