WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED HELICAL BLADE 100MM - STERILE; ROD, FIXATION, INTRAMEDULLARY
|
Back to Search Results |
|
| Model Number |
04.038.400S |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020, during an orthopedic procedure, the surgeon experienced complications while using the proximal femoral nailing system (tfna) devices.Difficulties were encountered while using the stepped drill bit in preparation for the helical blade.After drilling, the surgeon proceeded to insert the helical blade component, which was difficult to insert.The surgeon proceeded to insert the distal locking screw and had complications with inserting it.Afterwards the surgeon noticed that an incorrect aiming arm (125 degree) was utilized.The tfna nail that was implanted was a 130 degree nail.It is unknown if there was a surgical delay.Patient status and surgical outcome were unknown.Concomitant device reported: tfna nail (part 04.037.142s, lot unknown, quantity 1), cannulated stepped drill bit (part 03.037.022, lot unknown, quantity 1), three-fluted drill bit (part 03.010.061, lot unknown, quantity 1), 5.0mm ti locking screw w/t25 stardrive 42mm f/im nail-ster (part 04.005.532s, lot unknown, quantity 1), guides/sleeves/aiming: aiming arm (part number unknown, lot unknown, quantity 1).This report involves one tfna fenestrated helical blade 100mm ¿ sterile.This is report 1 of 3 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|