Concomitant medical products: plc271000/14082238, udi: (b)(4).Of the contralateral leg components implanted on july 21, 2016, it is unknown which component was implanted on which side.Please reference mfr report #2953161-2020-00010.
|
|
On (b)(6) 2016, this patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.On (b)(6) 2020, bilateral type ib endoleaks were discovered.The endoleaks were caused by bilateral degenerative iliac disease.On (b)(6) 2020, a reintervention occurred to treat the endoleaks.On the left side, coil embolization of the left internal iliac artery was performed along with an implant of a contralateral leg component to extend distally to the left external iliac artery.On the right side, a retrograde implant of an ibe device was performed (off label) to extend and bridge the previously implanted contralateral leg component.
|