| Model Number |
7135 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The target lesion was located in a left coronary artery.After a 300cm non-bsc support guidewire crossed the lesion, a 4.00mm x 30mm emerge balloon catheter was advanced for dilatation.However, during withdrawal, the balloon could not move without moving the wire.The procedure was completed with a different device.No patient complications nor injuries were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment occurred.The target lesion was located in a left coronary artery.After a 300cm non-bsc support guidewire crossed the lesion, a 4.00mm x 30mm emerge balloon catheter was advanced for dilatation.However, during withdrawal, the balloon could not move without moving the wire.The procedure was completed with a different device.No patient complications nor injuries were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 119cm distal of the strain relief.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.There were pinholes 1mm proximal of the proximal marker band, at the proximal marker band, and 2mm, 4mm, 6mm, and 8 mm distal of the proximal marker band.The inner shaft was buckled 14cm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Search Alerts/Recalls
|