| Model Number |
IPN000073 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the balloon tip of the catheter is twisted.The device is placed at the jugularis and used for ballooning of the vessel for dogs (veterinary use).When the balloon tip is twisted it is very difficult to move the catheter in the vessel causing a delay resulting in an extension of the anesthesia." there was no report of harm or complications to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn#(b)(4).The reported complaint that the "balloon tip of the catheter is twisted" is not able to be confirmed.A representative device was received for investigation with no damage or abnormalities noted.The balloon inflated as per specifications during the functional investigation.The returned device passed visual and functional test specifications.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported that "the balloon tip of the catheter is twisted.The device is placed at the jugularis and used for ballooning of the vessel for dogs (veterinary use).When the balloon tip is twisted it is very difficult to move the catheter in the vessel causing a delay resulting in an extension of the anesthesia." there was no report of harm or complications to the patient.
|
| |
|
Search Alerts/Recalls
|