DJO, LLC CMF SPINALOGIC; STIMULATOR, BONE GROWTH, NON-INVASIVE Back to Search Results
Model Number 01-207-0007
Medical Device Problem Codes Fluid/Blood Leak (1250); Battery Problem (2885)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/17/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device was returned for evaluation.If the device is received, a follow-up report will be submitted upon completion of product evaluation.
 
Event or Problem Description
It was reported that the patient "used the device since the beginning of january and when she tried it today she heard a popping sound and now the battery is leaking." no further information is currently available.
 
Additional Manufacturer Narrative
G1, g2: updated manufacturer contact.H2 - h6: device was returned for product evaluation.Device was returned with a swollen battery which did not properly fit within the control box.Device was tested with a known battery in good condition and results within device specifications were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMF SPINALOGIC
Common Device Name
STIMULATOR, BONE GROWTH, NON-INVASIVE
Manufacturer (Section D)
DJO, LLC
1430 decision street
vista, ca
MDR Report Key9891706
Report Number3012446970-2020-00012
Device Sequence Number13972642
Product Code LOF
UDI-Device Identifier00888912011495
UDI-Public00888912011495
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P910066
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number01-207-0007
Device Catalogue Number01-207-0007
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2020
Supplement Date Received by Manufacturer03/17/2020
Initial Report FDA Received Date03/27/2020
Supplement Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-