| Model Number |
7110-0308-3D-A2 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Stretched (1601); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Neurological Deficit/Dysfunction (1982); Perforation of Vessels (2135)
|
| Date of Event |
02/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was not returned to the manufacturer for evaluation.The root cause cannot be determined.The coil detachment occurred after the aneurysm perforation; therefore, the device malfunction did not cause or contribute to the adverse event.Additionally, it is unknown if the coil or the microcatheter (not a microvention device) perforated the aneurysm.
|
| |
|
Event or Problem Description
|
|
It was reported that treatment was performed for a choroidal aneurysm.During placement of the embolization coil, the aneurysm was perforated.During the attempt to retract the coil, the coil stretched and then detached.Another coil was placed in the aneurysm, which was successful in stopping the extravasation.The detached coil segment was then removed from the patient with a stent retriever.The patient woke up from anesthesia.The patient was reported to have had a stroke, but was slowly improving.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received from the sales representative indicated that the patient presented to the hospital with a subarachnoid hemorrhage.
|
| |
|