|
The reported event is inconclusive.No sample was returned for evaluation.It is therefore unknown if the device failed to meet specifications.The device was being used for treatment.A potential root cause of the reported event could be "outside diameter is too large".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿released bard® intermittent catheters (latex) manufacturer: c.R.Bard, inc.Covington, ga 30014 1-800-526-4455 www.Bardmedical.Com assembled in mexico instructions for use indications for use: for urological use only.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: this product contains natural rubber latex which may cause allergic reactions.Warnings: ¿ on latex and red rubber catheters, do not use ointments or lubricants having a petrolatum base.They will damage latex.¿ visually inspect the product for any imperfections or surface deterioration prior to use.¿ reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to failure, and/or to injury, illness or death of the patient.Note: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Adverse reactions urinary tract infection bleeding from the urethra irritation of the urethra instructions for intermittent catheters: 1.Wash your hands thoroughly with soap and water.2.Remove catheter from the pack.3.Position yourself comfortably, cleaning the opening of the urethra and surrounding area.4.Gently insert rounded end of catheter into urethra.5.When urine stops flowing, remove catheter from urethra.6.Dispose of catheter in accordance with local rules and regulations.7.Wash your hands.Caution: federal (u.S.A.) laws restricts this device to sale by or on the order of a physician.Single-use.Single patient use, do not reuse.Do not use if package is opened or damaged.Bard, bardex and bardia are trademarks and/or registered trademarks of c.R.Bard, inc.Pk7639306 06/2015 © 2015.C.R.Bard, inc.".
|