| Catalog Number |
55840006545 |
| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient presented with deformity; and underwent a revision of spine surgery at l3-s1.Intra-op, while trying to implant the alleged screw at sacrum, it was found that the screw was unable to turn further.The surgeon wasn't able to go through the second cortex.The screw was taken out; and it was found that the thread, at the distal part of the screw, came loose.The alleged screw was completely explanted.It appeared that while the screw was in the sacrum, it might have touched another screw (some part) which was still there.This might be a reason for the screw thread to have loosened; but the surgeon was not sure.No patient complications were reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: product analysis: visual and optical inspection confirmed the screw was returned damaged.The head of the screw has not been damaged and is still able to pivot freely.About 1 mm of the thread lip of the screw threads has been shaved off.This appears to be the result of metal on metal contact such as a screw or implant while tightening.Additional information: d10, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|