| Model Number |
B1100-080 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Event date is estimated.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other complaints.Based on the information provided, the reported balloon rupture was likely due to circumstances of the procedure.It is likely that the difficulty advancing, and balloon rupture were the result of interaction with the previously placed stents.Additionally, it is likely that during removal, the ruptured balloon material caught on the introducer sheath, resulting in difficulty removing and separation of the balloon.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Attachment: medsun report number: 2300200000-2020-8002.
|
| |
|
Event or Problem Description
|
|
User facility medwatch report received that states: "patient brought to ir, accompanied by crna and rn, for an outpatient scheduled fistulogram with possible intervention.An angioplasty was attempted.The balloon used burst and became lodged in patient.After unsuccessful attempts were made to retrieve balloon, the plan was for the patient to be taken to or to have balloon removed surgically.Patient was removed from procedure table and taken to post op.After consulting additional physicians, the decision was made to bring the patient back to ir for a different approach.The patient returned to ir accompanied by crna and rn.After changing the approach the femoral access, the balloon was retrieved using a snare.When questioning staff, it was noted that there were stents in the patient that might have snagged the balloon, however it is unclear what the root cause was.Unfortunately, due to the condition of the balloon, staff discarded the device post-op.There was no harm to the patient, and they were discharged same day." no additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being resubmitted to ensure the enclosed attachment can be easily opened by the fda.
|
| |
|