• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT X3 ELEVATED RIM 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT X3 ELEVATED RIM 36MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 643-00-36F
Medical Device Problem Codes Degraded (1153); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 02/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported that a revision surgery is scheduled for the patient's right hip on (b)(6) 2020 due to the head and part of the stem breaking off from the rest of the stem.Update: "surgeon revised liner, do to wear and time frame the cup had been in patient.".
 
Event or Problem Description
It was reported that a revision surgery is scheduled for the patient's right hip on (b)(6) 2020 due to the head and part of the stem breaking off from the rest of the stem.Update: "surgeon revised liner, do to wear and time frame the cup had been in patient.".
 
Additional Manufacturer Narrative
An event regarding wear involving trident liner was reported.The event was confirmed on the basis of visual inspection.Method & results: -device evaluation and results: visual inspection - as per the attached picture of explanted liner in communication log, liner appears to be in damaged state with various scratches in the inner surface and blackening at its outer circular surface area.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant rejected and stated the following comment: rejected for medical review - x-ray confirms disassociation and fracture at base of the trunnion of the stem, need additional information; revision operative reports, clinical and past medical history, additional serial dated x-rays and examination of explanted components.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT X3 ELEVATED RIM 36MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9898067
Report Number0002249697-2020-00631
Device Sequence Number2818518
Product Code LPH
UDI-Device Identifier07613327025101
UDI-Public07613327025101
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2013
Device Model Number643-00-36F
Device Catalogue Number643-00-36F
Device Lot NumberMER4XE
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/02/2020
Supplement Date Received by Manufacturer03/31/2020
Initial Report FDA Received Date03/30/2020
Supplement Report FDA Received Date04/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age80 YR
Patient Weight98
-
-