| Model Number |
643-00-36F |
| Medical Device Problem Codes |
Degraded (1153); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
02/11/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery is scheduled for the patient's right hip on (b)(6) 2020 due to the head and part of the stem breaking off from the rest of the stem.Update: "surgeon revised liner, do to wear and time frame the cup had been in patient.".
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery is scheduled for the patient's right hip on (b)(6) 2020 due to the head and part of the stem breaking off from the rest of the stem.Update: "surgeon revised liner, do to wear and time frame the cup had been in patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding wear involving trident liner was reported.The event was confirmed on the basis of visual inspection.Method & results: -device evaluation and results: visual inspection - as per the attached picture of explanted liner in communication log, liner appears to be in damaged state with various scratches in the inner surface and blackening at its outer circular surface area.Material analysis, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant rejected and stated the following comment: rejected for medical review - x-ray confirms disassociation and fracture at base of the trunnion of the stem, need additional information; revision operative reports, clinical and past medical history, additional serial dated x-rays and examination of explanted components.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|