Model Number 3L93712
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Atrial Fibrillation (1729); Foreign Body Reaction (1868); Rash (2033); Weakness (2145); Anxiety (2328); Discomfort (2330); Polydipsia (2604); No Code Available (3191)
Date of Event 03/16/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Litigation alleges that patient experienced pain, weakness, difficulty with simple daily living activities, inability to return to work, and increased metallic ions in his bloodstream.Revision operative note reviewed and confirmed metallosis.The revision operative note also mentions alval.In addition to what previously alleged, litigation alleged metal wear, skin rashes, weight loss, pain, increase incidence of atrial fibrillation, anxiety, thirst, loose of taste, metallic taste in saliva, and dry fragile skin.After review of medical records it was indicated that the patient present with progressive worsening left hip pain, demonstrated elevated metal ions.The patient was then revised for painful metal-sensitivity reaction to necrosis (alval).Operative notes reported that the cup was little generously anteverted and there was some erosive deficiency noted of the anterior acetabular bony wall, due to inflammatory process.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL2 LAT COXA VARA SIZE 12
Common Device Name
CORAIL AMT CEMENTLESS IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
Manufacturer (Section G)
DEPUY FRANCE SAS 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR   69801
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
b.p. 256
warsaw 46581
6107428552
MDR Report Key9899654
Report Number1818910-2020-09613
Device Sequence Number14144507
Product Code KWA
UDI-Device Identifier10603295168898
UDI-Public10603295168898
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K042992
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/01/2013
Device Model Number3L93712
Device Catalogue Number3L93712
Device Lot Number2743913
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/15/2020
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date03/30/2020
Supplement Report FDA Received Date02/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/31/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight48 KG
-
-