• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PINNACLE 300 ACET CUP 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PINNACLE 300 ACET CUP 50MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-03-050
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Atrial Fibrillation (1729); Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Rash (2033); Weakness (2145); Anxiety (2328); Discomfort (2330); Polydipsia (2604); Test Result (2695); No Code Available (3191)
Date of Event 03/16/2011
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).  if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Litigation alleges that patient experienced pain, weakness, difficulty with simple daily living activities, inability to return to work, and increased metallic ions in his bloodstream.Revision operative note reviewed and confirmed metallosis.The revision operative note also mentions alval.In addition to what previously alleged, litigation alleged metal wear, skin rashes, weight loss, pain, increase incidence of atrial fibrillation, anxiety, thirst, loose of taste, metallic taste in saliva, and dry fragile skin.After review of medical records it was indicated that the patient present with progressive worsening left hip pain, demonstrated elevated metal ions.The patient was then revised for painful metal-sensitivity reaction to necrosis (alval).Operative notes reported that the cup was little generously anteverted and there was some erosive deficiency noted of the anterior acetabular bony wall, due to inflammatory process.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 300 ACET CUP 50MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic drive
warsaw, IN 46581-0988
6107428552
MDR Report Key9899671
Report Number1818910-2020-09610
Device Sequence Number19539910
Product Code LPH
UDI-Device Identifier10603295008644
UDI-Public10603295008644
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K000306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/23/2012
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-03-050
Device Catalogue Number121703050
Device Lot NumberDG4F21000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 03/15/2020
Supplement Date Received by Manufacturer02/18/2022
Initial Report FDA Received Date03/30/2020
Supplement Report FDA Received Date02/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/27/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight48 KG
-
-