COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Model Number 88-30414001 |
Device Problem
Product Quality Problem (1506)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, post operatively, after inserting the catheter via right jugular position, it was stated that it was secured in place by binding it to a plastic ring inserted/secured in the patient skin, 2 days after, it was noted that the catheter slipped out of place without any external force applied.Due to the slipping, the catheter was removed.There was no reported patient consequence.
|
|
Manufacturer Narrative
|
Additional info: g4, h3, h6 h3.Evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.The visual inspection of the returned photos noted: the first image depicts a gloved hand separating the blue and red luer adapters.There is blood in the extension tubes and an unidentifiable substance on the hub.The second image depicts the suture wing, ring and gauze on a patient.Without the physical device functional evaluations are precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|