• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK; PLATE, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source ¿ (b)(6).
 
Event or Problem Description
It was reported by the surgeon that the grip was extremely weak compared to other screws.The event occurred during the surgery.No adverse events have been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The device was returned for investigation.The screw was visually evaluated and found to show signs of use.The screw shows signs of use/ wear at the cross-drive interface on the head of the screw.Functional testing was done for retention using a driver 01-7390 lot 048760 and bit 15-1196 lot 406650.The blade was inserted into the driver and the screw, then the assembly was lifted by the screw to verify the cross-drive feature could support the weight of the blade and driver.The screw failed this retention test.The dhr will not be reviewed as the lot number remains unknown.There are no indications of manufacturing defects.For this part (95-6104) and the previous one year (from the notification date) regarding retention failure, (b)(4).The most likely underlying cause of the complaint cannot be determined.However, potential contributing factors are using a worn or incorrect blade and incorrect alignment of the blade and screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK
Common Device Name
PLATE, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9900444
Report Number0001032347-2020-00165
Device Sequence Number16835348
Product Code JEY
UDI-Device Identifier00841036056650
UDI-Public00841036056650
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121589
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number95-6104
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/12/2020
Supplement Date Received by Manufacturer09/17/2020
Initial Report FDA Received Date03/30/2020
Supplement Report FDA Received Date09/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-