| Model Number |
N/A |
| Medical Device Problem Code |
Device-Device Incompatibility (2919)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source ¿ (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported by the surgeon that the grip was extremely weak compared to other screws.The event occurred during the surgery.No adverse events have been reported as a result of the malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The complaint is confirmed.The device was returned for investigation.The screw was visually evaluated and found to show signs of use.The screw shows signs of use/ wear at the cross-drive interface on the head of the screw.Functional testing was done for retention using a driver 01-7390 lot 048760 and bit 15-1196 lot 406650.The blade was inserted into the driver and the screw, then the assembly was lifted by the screw to verify the cross-drive feature could support the weight of the blade and driver.The screw failed this retention test.The dhr will not be reviewed as the lot number remains unknown.There are no indications of manufacturing defects.For this part (95-6104) and the previous one year (from the notification date) regarding retention failure, (b)(4).The most likely underlying cause of the complaint cannot be determined.However, potential contributing factors are using a worn or incorrect blade and incorrect alignment of the blade and screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|