• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER SIZE 11 STANDARD OFFSET; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. FEMORAL STEM 12/14 NECK TAPER SIZE 11 STANDARD OFFSET; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Material Erosion (1214); Insufficient Information (3190)
Patient Problems Host-Tissue Reaction (1297); Pain (1994); No Information (3190)
Event Date 09/18/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical product(s): 00801803602 femoral head 61357236; 00875101336 liner neutral 36 mm 61303880; 00875705802 shell with multi holes 61321986; 00625006530 bone scr 61341911; 00625006535 bone scr 61341914.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0002648920 - 2020 - 00209 head.
 
Event Description
It was reported that the patient underwent left total hip arthroplasty and was later revised nine years later due to unknown reasons.The femoral head was exchanged.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(6).Reported event was confirmed by review of medical records.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.Medical records identified the following : patient underwent a revision procedure on due to elevated metal ions, pain, corrosion, and pseudotumor.Blood work showed that patient had elevated meta ions.Post revision blood work showed a reduction in the metal ion levels during the revision procedure, corrosion was noted at the head/ neck junction.Pseudotumor was observed in the joint space.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4) this follow-up report is being submitted to relay additional information.Updated: a1, a2, a6, b4, b5, b6, g4, h2, h3, h6 if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the patient underwent left total hip arthroplasty and was later revised nine years later due to left hip pain, elevated metal ions, corrosion and pseudotumor.Attempts have been made and no further information has been provided.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Updated: a4, b4, b5, b7, g4, h2 corrected: e1 if any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Reported event was confirmed by review of medical records.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.The received additional information does not change the final results of previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL STEM 12/14 NECK TAPER SIZE 11 STANDARD OFFSET
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9900801
MDR Text Key191866121
Report Number0001822565-2020-01113
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
K161830
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,health p
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 08/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Model NumberN/A
Device Catalogue Number00771101100
Device Lot Number61273064
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight107
-
-