| Model Number |
383074 |
| Medical Device Problem Code |
Failure to Capture (1081)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the his bundle lead was not capturing.The lead was explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the full lead was returned and analyzed.No anomalies were found.The distal low voltage electrode of the lead was covered in body tissue.Visual analysis of the lead indicated apparent explant damage.If information is provided in the future, a supplemental report will be issued.
|
| |
|