| Catalog Number |
1553201080 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that on: (b)(6) 2018: the patient underwent a spinal fusion.In (b)(6) 2019: the patient underwent x-ray/mri, which showed the rods on both sides to have broken.Hence, the patient was scheduled for rod removal in (b)(6) 2020.On (b)(6) 2020: the patient presented with the pre-operative diagnoses of broken rods and pseudoarthrosis.She underwent posterior lumbar fusion with transforaminal lumbar interbody fusion at l3-s1.The broken rods were returned to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: x-ray review results: post-op x-ray from l4-s1 posterior spinal instrumentation is provided.No date is on the image.Both of the rods have fractured between l4 and l5.No interbody device is present.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|