Model Number 1871
Device Problems Detachment of Device or Device Component (2907); Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/10/2020
Event Type  malfunction  
Event Description
It was reported that the collar of the device was cracked.The target lesion was in the coronary artery.A 3.00x38 synergy drug eluting stent and a guidezilla ii guide extension catheter were selected for use in a percutaneous coronary intervention (pci).During the procedure, it was noted that the stent was stripped off from the balloon at the collar of the guidezilla.The stent delivery system was then removed to the collar of the guidezilla and were removed together.However, a crack at the collar of the guidezilla was noted after removal.There were no device fragments left inside the patient's body and the procedure was completed with a different device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDEZILLA II
Type of Device
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key9903235
MDR Text Key185943942
Report Number2134265-2020-04100
Device Sequence Number1
Product Code DQY
UDI-Device Identifier08714729939450
UDI-Public08714729939450
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163314
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/06/2021
Device Model Number1871
Device Catalogue Number1871
Device Lot Number0024558658
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/08/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-