MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE Back to Search Results
Model Number M00561311
Medical Device Problem Codes Use of Device Problem (1670); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Burn(s) (1757)
Date of Event 01/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a captivator large oval med stiff snare was used in the colon during a polypectomy procedure performed on (b)(6) 2020.According to the complainant, during procedure and outside the patient, the snare was passed several times in and out of the scope and at some point, the catheter got detached from the handle leaving the wire at the top exposed.When the snare was removed, the exposed wire was not noticed and was placed externally on the patient.While the endoscopist was removing the scope, he stepped on the diathermy pedal and the snare burned the patient's leg.The procedure was completed with this device.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR
Common Device Name
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9903894
Report Number3005099803-2020-01412
Device Sequence Number13616704
Product Code FDI
UDI-Device Identifier08714729019251
UDI-Public08714729019251
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K131700
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/24/2021
Device Model NumberM00561311
Device Catalogue Number6131
Device Lot Number0022297845
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2020
Initial Report FDA Received Date03/31/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-