| Model Number |
1012450-15 |
| Medical Device Problem Codes |
Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/11/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a mildly calcified, mildly tortuous de novo left anterior descending artery that was 80% stenosed.The 3.25x15mm nc trek balloon was inflated a few times (exact number not given) at 20 atmospheres (atm) then the balloon ruptured at 20 atm.A new unspecified balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the returned device.The balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use states: balloon pressure should not exceed the rated burst pressure (rbp).The investigation determined the reported balloon rupture appears to be related to the violation of the ifu as the nc trek bdc was over inflated to 20 atmospheres (atms) multiple times.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|