Model Number 1012450-15
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/11/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.(b)(4).
 
Event or Problem Description
It was reported that the procedure was performed to treat a mildly calcified, mildly tortuous de novo left anterior descending artery that was 80% stenosed.The 3.25x15mm nc trek balloon was inflated a few times (exact number not given) at 20 atmospheres (atm) then the balloon ruptured at 20 atm.A new unspecified balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned device.The balloon rupture was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use states: balloon pressure should not exceed the rated burst pressure (rbp).The investigation determined the reported balloon rupture appears to be related to the violation of the ifu as the nc trek bdc was over inflated to 20 atmospheres (atms) multiple times.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX & VOYAGER NC
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9904437
Report Number2024168-2020-03140
Device Sequence Number10938218
Product Code LOX
UDI-Device Identifier08717648151934
UDI-Public08717648151934
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model Number1012450-15
Device Catalogue Number1012450-15
Device Lot Number90928G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2020
Initial Date Received by Manufacturer 03/12/2020
Supplement Date Received by Manufacturer04/14/2020
Initial Report FDA Received Date03/31/2020
Supplement Report FDA Received Date04/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-